Abstract
Laser resonance ionization at the RISIKO 30 kV mass separator has been used to produce isotopically and isobarically pure and well quantified 222Rn emanation standards. Based upon laser-spectroscopic preparation studies, ion implantation into aluminum and tungsten targets has been carried out, providing overall implantation efficiencies of 40% up to 60%. The absolute implanted activity of 226Ra was determined by the technique of defined solid-angle α-particle spectrometry, where excellent energy resolution was observed. The 222Rn emanation coefficient of the produced targets was studied using α-particle and γ-ray spectrometry, and yielded results between 0.23 and 0.34, with relative uncertainty on the order of 1%. No dependence exceeding a 1% change of the emanation on humidity could be identified in the range of 15 %rH to 75 %rH, whereas there were hints of a slight correlation between the emanation and temperature. Additionally, and as expected, the emanation coefficient was found to be dependent on the target material as well as the implanted dose.
| Originalsprache | Englisch |
|---|---|
| Aufsatznummer | 110093 |
| Fachzeitschrift | Applied radiation and isotopes |
| Jahrgang | 181 |
| Elektronisch veröffentlicht (E-Pub) | 31 Dez. 2021 |
| DOIs | |
| Publikationsstatus | Veröffentlicht - März 2022 |
ASJC Scopus Sachgebiete
- Strahlung
Dieses zitieren
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver