A simple light path modifying device to reduce scattering in front-face fluorescence spectra

Frank B. Peters, Andreas Otto Rapp

Research output: Contribution to journalArticleTransferpeer review

Abstract

This technical note presents a device to diminish scattering signal in front-face fluorescence spectra while obtaining fluorescence signal. The beam path in a commercial fluorescence spectrometer was modified by two deflecting mirrors, leading reflections away from the sensor. This light path modifying (LPM) device was tested with two fluid and three solid substances, where the scattering-to-fluorescence ratio improved by a factor of 1.7 to 7.6. The spectra obtained with the LPM were much clearer, and distortion of the fluorescence peaks was avoided. Scans of quinine sulphate complied well with reference spectra.
Original languageEnglish
Article number037001
JournalMethods and Applications in Fluorescence
Volume12
Issue number3
DOIs
Publication statusPublished - 17 Jun 2024