Clinical Results and Mechanical Properties of the Carotid CGUARD Double-Layered Embolic Prevention Stent

  • Christian Wissgott
  • , Wolfram Schmidt
  • , Christoph Brandt-Wunderlich
  • , Peter Behrens
  • , Reimer Andresen

Research output: Contribution to journalArticleResearch

Abstract

PURPOSE: To report early clinical outcomes with a novel double-layer stent for the internal carotid artery (ICA) and the in vitro investigation of the stent's mechanical properties.

METHODS: A prospective single-center study enrolled 30 consecutive patients (mean age 73.1±6.3 years; 21 men) with symptomatic (n=25) or high-grade (n=5) ICA stenosis treated with the new double-layer carotid CGUARD Embolic Prevention System (EPS) stent, which has an inner open-cell nitinol design with an outer closed-cell polyethylene terephthalate layer. The average stenosis of the treated arteries was 84.1%±7.9% with a mean lesion length of 16.6±2.1 mm. In the laboratory, 8×40-mm stents where tested in vitro with respect to their radial force during expansion, the bending stiffness of the stent system and the expanded stent, as well as the collapse pressure in a thin and flexible sheath. The wall adaptation was assessed using fluoroscopy after stent release in step and curved vessel models.

RESULTS: The stent was successfully implanted in all patients. No peri- or postprocedural complications occurred; no minor or major stroke was observed in the 6-month follow-up. The bending stiffness of the expanded stent was 63.1 N·mm2 and (not unexpectedly) was clearly lower than that of the stent system (601.5 N·mm2). The normalized radial force during expansion of the stent to 7.0 mm, consistent with in vivo sizing, was relatively high (0.056 N/mm), which correlates well with the collapse pressure of 0.17 bars. Vessel wall adaptation was harmonic and caused no straightening of the vessel after clinical application.

CONCLUSION: Because of its structure, the novel CGUARD EPS stent is characterized by a high flexibility combined with a high radial force and very good plaque coverage. These first clinical results demonstrate a very safe implantation behavior without any stroke up to 6 months after the procedure.

Original languageEnglish
Pages (from-to)130-137
Number of pages8
JournalJournal of Endovascular Therapy
Volume24
Issue number1
DOIs
Publication statusPublished - Feb 2017
Externally publishedYes

Keywords

  • Aged
  • Angiography, Digital Subtraction
  • Angioplasty, Balloon/adverse effects
  • Carotid Artery, Internal/diagnostic imaging
  • Carotid Stenosis/diagnostic imaging
  • Embolic Protection Devices
  • Female
  • Germany
  • Humans
  • Male
  • Materials Testing
  • Prospective Studies
  • Prosthesis Design
  • Prosthesis Failure
  • Severity of Illness Index
  • Stents
  • Time Factors
  • Treatment Outcome
  • open-cell design
  • stenosis
  • mechanical behavior
  • radial force
  • internal carotid artery
  • closed-cell design
  • stent
  • embolic filter
  • in vitro testing
  • double-layer stent
  • nitinol
  • carotid artery stent