Projects per year
Abstract
In this work, the influence of the sample gas flow rate and the ionization region volume of an ion mobility spectrometer (IMS) used as a detector in gas chromatography (GC) on GC-IMS peak shape has been investigated. Therefore, a drift tube IMS with a field-switching ion shutter, a defined ionization region volume and an ultra-violet radiation source was used. To identify the influence of the sample gas flow rate entering the ionization region (equals the GC carrier gas flow rate if no further make-up gas is used) and the ionization region volume on peak broadening and signal intensity, different sample volumes as they would elute from a GC were tested at a variety of sample gas flow rates at a given ionization region volume. The results clearly show that for low sample gas flow rates a depletion of sample molecules in the ionization region leads to a significant decrease in effective detector volume but also to reduced signal intensities. Therefore, for optimal performance of a GC-IMS, the optimal operating point of the GC should match the flow range, where the IMS provides the best compromise between signal-to-noise ratio and peak broadening.
| Original language | English |
|---|---|
| Article number | 464506 |
| Number of pages | 8 |
| Journal | Journal of Chromatography A |
| Volume | 1713 |
| E-pub ahead of print | 13 Nov 2023 |
| DOIs | |
| Publication status | Published - 4 Jan 2024 |
Keywords
- Effective detector volume
- GC-IMS
- Ionization volume
- Peak broadening
ASJC Scopus subject areas
- Analytical Chemistry
- Biochemistry
- Organic Chemistry
Projects
- 1 Finished
-
SIIRI: Collaborative Research Centre/Transregio 298/1: Safety Integrated and Infection Reactive Implants
Stiesch, M. (Principal Investigator) & Maier, H. J. (Co-Principal Investigator)
1 Jul 2021 → 31 Dec 2025
Project: Research
Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver