Localization accuracy of sphere fiducials in computed tomography images

  • Jan Philipp Kobler*
  • , Jesús Díaz Díaz
  • , J. Michael Fitzpatrick
  • , G. Jakob Lexow
  • , Omid Majdani
  • , Tobias Ortmaier
  • *Corresponding author for this work

Research output: Chapter in book/report/conference proceedingConference contributionResearchpeer review

Abstract

In recent years, bone-attached robots and microstereotactic frames have attracted increasing interest due to the promising targeting accuracy they provide. Such devices attach to a patient's skull via bone anchors, which are used as landmarks during intervention planning as well. However, as simulation results reveal, the performance of such mechanisms is limited by errors occurring during the localization of their bone anchors in preoperatively acquired computed tomography images. Therefore, it is desirable to identify the most suitable fiducials as well as the most accurate method for fiducial localization. We present experimental results of a study focusing on the fiducial localization error (FLE) of spheres. Two phantoms equipped with fiducials made from ferromagnetic steel and titanium, respectively, are used to compare two clinically available imaging modalities (multi-slice CT (MSCT) and cone-beam CT (CBCT)), three localization algorithms as well as two methods for approximating the FLE. Furthermore, the impact of cubic interpolation applied to the images is investigated. Results reveal that, generally, the achievable localization accuracy in CBCT image data is significantly higher compared to MSCT imaging. The lowest FLEs (approx. 40 μm) are obtained using spheres made from titanium, CBCT imaging, template matching based on cross correlation for localization, and interpolating the images by a factor of sixteen. Nevertheless, the achievable localization accuracy of spheres made from steel is only slightly inferior. The outcomes of the presented study will be valuable considering the optimization of future microstereotactic frame prototypes as well as the operative workflow.

Original languageEnglish
Title of host publicationMedical Imaging 2014
Subtitle of host publicationImage-Guided Procedures, Robotic Interventions, and Modeling
PublisherSPIE
ISBN (Print)9780819498298
DOIs
Publication statusPublished - 2014
EventMedical Imaging 2014: Image-Guided Procedures, Robotic Interventions, and Modeling - San Diego, CA, United States
Duration: 18 Feb 201420 Feb 2014

Publication series

NameProgress in Biomedical Optics and Imaging - Proceedings of SPIE
Volume9036
ISSN (Print)1605-7422

Conference

ConferenceMedical Imaging 2014: Image-Guided Procedures, Robotic Interventions, and Modeling
Country/TerritoryUnited States
CitySan Diego, CA
Period18 Feb 201420 Feb 2014

Keywords

  • accuracy
  • fiducial localization error
  • medical image processing
  • surgical robotics

ASJC Scopus subject areas

  • Electronic, Optical and Magnetic Materials
  • Atomic and Molecular Physics, and Optics
  • Biomaterials
  • Radiology Nuclear Medicine and imaging

Cite this