Skip to main navigation Skip to search Skip to main content

System Design and Optimization of a Miniaturized Ion Mobility Spectrometer Using Finite-Element Analysis

Research output: Contribution to journalArticleResearchpeer review

Abstract

Ion mobility spectrometry (IMS) is a widely used method for detecting low ppb-level concentrations of hazardous compounds, such as chemical warfare agents, explosives, and pollutants in air. Simple instrumentation combined with high sensitivity and high selectivity are major advantages of this technique. We are currently developing a miniaturized low-cost IMS that can be used in various applications, where fast detection of chemical compounds at low ppb-levels is required and space is limited. A numerical model based on finite-element analysis has been developed for rapid system design and optimization with respect to physical and economic constraints. All significant physical effects on the ion transport in an IMS are considered within the simulation, such as diffusion, coulomb repulsion, electric fields and gas flows. This model is a powerful tool for developing, studying, and optimizing miscellaneous types of IMS.

Original languageEnglish
Pages (from-to)377-382
Number of pages6
JournalIEEE Sensors Journal
Volume9
Issue number4
DOIs
Publication statusPublished - Apr 2009
Externally publishedYes

Keywords

  • Coulomb repulsion
  • finite-element analysis
  • ion mobility spectrometry (IMS)
  • simulation
  • spectrometers

ASJC Scopus subject areas

  • Instrumentation
  • Electrical and Electronic Engineering

Cite this